Medrad Avanta
Technical Specifications*
Flow Rates Contrast Fixed Flow: User defined from 1–45 mL/ sec in increments of 1 mL/sec; Variable Flow: User controlled variable range from 1 to 10 mL/sec in 0.1 increments; Saline Fixed Rate: 1.0 mL/sec
Volume User controlled variable range 1–150 mL in increments of 1 mL
Pressure Limits User defined from 300 PSI to 1200 PSI in increments of 1 PSI
Fill Rate Manual or automatic refill 2 mL/sec
Syringe Maintainer 35°C ±5° (≥1 hr)
Injection Delay 0.1 sec to 0-99.9 sec
Rise Time Programmable at 0.1 sec increments from 0.1 to 9.9 sec
Memory Up to 40 customizable protocols store
Display Screen 26 cm (~10.5") Color Touch Screen
Mounting Options Pedestal or Table Mount
Power Requirements Nominal 100–120 VAC, 50 or 60 Hz, 12.75 A; 200–240 VAC, 50 or 60 Hz, 5.3A
ISI Capable Provides for synchronization of the injection and x-ray exposure
Air Management Gross Air Detection and Fluid Level Sensing
Hemodynamic Monitoring Automatic transition to hemodynamic waveform after saline and contrast injection (continuous invasive blood pressure monitoring)